Myths and maths in MR safety 3
December 14, 2021
Why the value of magnetic susceptibility doesn’t really matter #MRI #MRIsafety … More Myths and maths in MR safety 3
Why the value of magnetic susceptibility doesn’t really matter #MRI #MRIsafety … More Myths and maths in MR safety 3
The following lists a number of corrections to the current published version. Table 2.1: Ferromagnetic M is parallel to external H and B0 M is actually anti-parallel to the internal H, the “demagnetising field”. Figure 2.21: The force at each end is T/l (torque divided by length). The equation stated in the caption is the … More Essentials of MRI Safety- corrections & clarifications